NI 5402/5406 Calibration Procedure
2.
Set the analog filter state by calling the niFgen Property Node and
selecting Output Attributes»Analog Filter Enabled.
LabVIEW Block Diagram
3.
Set the pre-amplifier attenuation by calling the niFgen Property Node
and selecting Calibration»Pre-Amplifier Attenuation.
LabVIEW Block Diagram
62
C/C++
Function Call
Call
niFgen_
SetAttribute
using the
ViBoolean
following parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_ATTR_ANALOG_
FILTER_ENABLED
value: The Analog Filter
Enable value for the
current iteration from
Table 7
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_PRE_AMPLIFIER_
ATTENUATION
value: The Pre-Amplifier
Attenuation value for the
current iteration from
Table 7
.
.
ni.com