National Instruments NI 5402 Calibration Procedure page 47

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
14. Commit the attribute values to the device by calling the niFgen
Commit VI.
LabVIEW Block Diagram
15. Wait 500 ms for the output to settle.
16. Measure the analog output voltage with the onboard calibration ADC
by calling the niFgen Read CAL ADC VI.
LabVIEW Block Diagram
47
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
C/C++
Function Call
Call
niFgen_Read
using the
CalADC
following parameters:
vi: The session handle
returned from
niFgen_InitExtCal
numberOfReads
ToAverage:
3
returnCalibratedValue:
VI_FALSE
calADCValue: Returns a
ViReal64 variable. The
variable passed by
reference through this
parameter receives the
voltage measured by the
onboard ADC. This value
is
cal ADC
, which
measurement 0
is used in step 30.
NI 5402/5406 Calibration Procedure
.
.