National Instruments NI 5402 Calibration Procedure page 12

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

LabVIEW Block Diagram
Table 2. Frequency Ranges for Verifying Oscillator Frequency Accuracy
Calibration Test Limits ±4.5 ppm
Low
9,999,955 Hz
NI 5402/5406 Calibration Procedure
4.
Initiate waveform generation by calling the niFgen Initiate
Generation VI.
5.
Use the frequency counter to measure the frequency output at the
CH 0 front panel connector of the NI 5402/5406. This value is the
measured frequency of the generated sine wave.
6.
Compare the measured frequency of the generated sine wave to the
calibration test limits and published specifications shown in Table 2.
High
10,000,045 Hz
7.
Abort waveform generation by calling the niFgen Abort
Generation VI.
LabVIEW Block Diagram
Published Specifications ±25 ppm
Low
9,999,750 Hz
12
C/C++ Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init
High
10,000,250 Hz
C/C++ Function Call
Call
niFgen_Abort
Generation
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init