Table 6. Values for Verifying Square Wave Gain and Offset Accuracy (Continued)
Analog Static
Iteration
Value
72
FFFF8000
73
7FFF
74
FFFF8000
75
7FFF
76
FFFF8000
77
7FFF
78
FFFF8000
79
7FFF
80
FFFF8000
81
7FFF
82
FFFF8000
83
7FFF
84
FFFF8000
85
7FFF
86
FFFF8000
87
7FFF
88
FFFF8000
89
7FFF
90
FFFF8000
91
7FFF
92
FFFF8000
93
7FFF
94
FFFF8000
95
7FFF
96
FFFF8000
NI 5402/5406 Calibration Procedure
Amplitude
DC Offset
(V
)
(V)
pp
0.5
–0.25
0.35
0.175
0.35
–0.175
0.25
0.125
0.25
–0.125
0.15
0.075
0.15
–0.075
0.1
0.05
0.1
–0.05
0.07
0.035
0.07
–0.035
0.05
0.025
0.05
–0.025
0.035
0.0175
0.035
–0.0175
0.025
0.0125
0.025
–0.0125
0.02
0.01
0.02
–0.01
0.015
0.0075
0.015
–0.0075
0.012
0.005
0.012
–0.005
0.012
0.002
0.012
–0.002
34
Ideal DC
Calibration
Output
Test Limits
(V)
(V)
–0.5
±0.00175
0.35
±0.001525
–0.35
±0.001525
0.25
±0.001375
–0.25
±0.001375
0.15
±0.001225
–0.15
±0.001225
0.1
±0.00115
–0.1
±0.00115
0.07
±0.001105
–0.07
±0.001105
0.05
±0.001075
–0.05
±0.001075
0.035
±0.001053
–0.035
±0.001053
0.025
±0.001038
–0.025
±0.001038
0.02
±0.00103
–0.02
±0.00103
0.015
±0.001022
–0.015
±0.001022
0.011
±0.001017
–0.011
±0.001017
0.008
±0.001014
–0.008
±0.001014
Published
Specification
(V)
±0.00575
±0.004625
±0.004625
±0.003875
±0.003875
±0.003125
±0.003125
±0.00275
±0.00275
±0.002525
±0.002525
±0.002375
±0.002375
±0.0022625
±0.0022625
±0.0021875
±0.0021875
±0.00215
±0.00215
±0.0021125
±0.0021125
±0.002085
±0.002085
±0.00207
±0.00207
ni.com