10. Close the instrument driver session and save the calibration date and
temperature by calling the niFgen Close Ext Cal VI.
C/C++
LabVIEW Block Diagram
Function Call
Call
niFgen_CloseExtCal
using the following parameters:
vi: The session handle returned from
.
niFgen_InitExtCal
action: If the external adjustment
procedure completed without any
errors, use
NIFGEN_VAL_EXT_CAL_
. This function then stores the
COMMIT
new calibration constants, updated
calibration dates, updated calibration
temperatures in the onboard
EEPROM.
If any errors occurred during the
external adjustment procedure, or if
you want to abort the operation, use
.
NIFGEN_VAL_EXT_CAL_ABORT
This function then discards the new
calibration constants and does not
change any of the calibration data
stored in the onboard EEPROM.