Table 6. Values for Verifying Square Wave Gain and Offset Accuracy (Continued)
Analog Static
Iteration
Value
14
FFFF8000
15
7FFF
16
FFFF8000
17
7FFF
18
FFFF8000
19
7FFF
20
FFFF8000
21
7FFF
22
FFFF8000
23
7FFF
24
FFFF8000
25
7FFF
26
FFFF8000
27
7FFF
28
FFFF8000
29
7FFF
30
FFFF8000
31
7FFF
32
FFFF8000
33
7FFF
34
FFFF8000
35
7FFF
36
FFFF8000
37
7FFF
38
FFFF8000
39
7FFF
40
FFFF8000
41
7FFF
42
FFFF8000
NI 5402/5406 Calibration Procedure
Amplitude
DC Offset
(V
)
(V)
pp
4
0
3
0
3
0
2.8
0
2.8
0
2
0
2
0
1.4
0
1.4
0
1
0
1
0
0.7
0
0.7
0
0.5
0
0.5
0
0.3
0
0.3
0
0.2
0
0.2
0
0.14
0
0.14
0
0.1
0
0.1
0
0.07
0
0.07
0
0.05
0
0.05
0
0.04
0
0.04
0
32
Ideal DC
Calibration
Output
Test Limits
(V)
(V)
–2
±0.005
1.5
±0.004
–1.5
±0.004
1.4
±0.0038
–1.4
±0.0038
1
±0.003
–1
±0.003
0.7
±0.0024
–0.7
±0.0024
0.5
±0.002
–0.5
±0.002
0.35
±0.0017
–0.35
±0.0017
0.25
±0.0015
–0.25
±0.0015
0.15
±0.0013
–0.15
±0.0013
0.1
±0.0012
–0.1
±0.0012
0.07
±0.00114
–0.07
±0.00114
0.05
±0.0011
–0.05
±0.0011
0.035
±0.00107
–0.035
±0.00107
0.025
±0.00105
–0.025
±0.00105
0.02
±0.00104
–0.02
±0.00104
Published
Specification
(V)
±0.022
±0.017
±0.017
±0.016
±0.016
±0.012
±0.012
±0.009
±0.009
±0.007
±0.007
±0.0055
±0.0055
±0.0045
±0.0045
±0.0035
±0.0035
±0.003
±0.003
±0.0027
±0.0027
±0.0025
±0.0025
±0.00235
±0.00235
±0.00225
±0.00225
±0.0022
±0.0022