National Instruments NI 5402 Calibration Procedure page 89

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

LabVIEW Block Diagram
© National Instruments Corporation
14. Close the instrument driver session and save the calibration date and
temperature by calling the niFgen Close Ext Cal VI.
Call
using the following
parameters:
vi: The session handle returned
from
action: If the external
adjustment procedure
completed without any errors,
use
COMMIT
stores the new calibration
constants, updated calibration
dates, updated calibration
temperatures in the onboard
EEPROM.
If any errors occurred during
the external adjustment
procedure, or if you want to
abort the operation, use
NIFGEN_VAL_EXT_CAL_
ABORT
discards the new calibration
constants and does not change
any of the calibration data
stored in the onboard
EEPROM.
89
C/C++
Function Call
niFgen_CloseExtCal
niFgen_InitExtCal
NIFGEN_VAL_EXT_CAL_
. This function then
. This function then
NI 5402/5406 Calibration Procedure
.