© National Instruments Corporation
10. Set the output impedance by calling the niFgen Property Node and
selecting Basic Operation»Output Impedance.
LabVIEW Block Diagram
11. Disable the analog output by calling the niFgen Property Node and
selecting Basic Operation»Output Enabled.
LabVIEW Block Diagram
45
C/C++ Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_ATTR_OUTPUT_
IMPEDANCE
value:
50.00
C/C++ Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_OUTPUT_ENABLED
value:
VI_FALSE