National Instruments NI 5402 Calibration Procedure page 53

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
32. Set the calibration ADC input by calling the niFgen Property Node and
selecting Calibration»Cal ADC Input.
LabVIEW Block Diagram
33. Commit the attribute values to the device by calling the niFgen
Commit VI.
LabVIEW Block Diagram
53
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
(empty string)
""
attributeID:
NIFGEN_
ATTR_CAL_ADC_INPUT
value:
NIFGEN_VAL_GROUND
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
NI 5402/5406 Calibration Procedure
.
.