NI 5402/5406 Calibration Procedure
c.
Commit the attribute values to the device by calling the niFgen
Commit VI.
LabVIEW Block Diagram
d.
Wait 500 ms for the output to settle.
e.
Use the DMM to measure the NI 5402/5406 output voltage. This
measurement is
f.
Set the gain DAC value by calling the niFgen Property Node and
selecting Calibration»Gain DAC Value.
LabVIEW Block Diagram
, which is used in step 5.
measurement 0
64
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_GAIN_DAC_VALUE
value:
1,000
.
.
ni.com