National Instruments NI 5402 Calibration Procedure page 65

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
g.
Commit the attribute values to the device by calling the niFgen
Commit VI.
LabVIEW Block Diagram
h.
Wait 500 ms for the output to settle.
i.
Use the DMM to measure the NI 5402/5406 output voltage. This
measurement is
j.
Set the offset DAC value by calling the niFgen Property Node and
selecting Calibration»Offset DAC Value.
LabVIEW Block Diagram
, which is used in step 5.
measurement 1
65
C/C++ Function Call
Call
niFgen_Commit
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_ATTR_OFFSET_
DAC_VALUE
value:
26,000
NI 5402/5406 Calibration Procedure
.