National Instruments NI 5402 Calibration Procedure page 87

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
a.
Set the frequency by calling the niFgen Property Node and
selecting Standard Function Output»Frequency.
LabVIEW Block Diagram
b.
Use the power meter to measure the power generated by the
NI 5402/5406 during the generation of the sine wave.
c.
Repeat steps 11a through 11b for each frequency in Table 11.
12. Abort waveform generation by calling the niFgen Abort
Generation VI.
LabVIEW Block Diagram
87
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_FUNC_FREQUENCY
value: The current
Frequency value from
Table 11
C/C++
Function Call
Call
niFgen_Abort
Generation
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
NI 5402/5406 Calibration Procedure
.
.