Table 5. Values for Verifying DC Gain and Offset Accuracy (Continued)
Iteration
Digital Gain
28
0
29
0
30
0
31
0
32
0
33
0
34
0
35
0
36
0
37
0
38
0
39
0
40
0
41
0
42
0
43
0
44
0
45
0
46
0
47
0
48
0
49
0
50
0
NI 5402/5406 Calibration Procedure
Amplitude
DC Offset
(V
)
(V)
pp
0.012
–0.269
0.012
0.194
0.012
–0.194
0.012
0.144
0.012
–0.144
0.012
0.094
0.012
–0.094
0.012
0.064
0.012
–0.064
0.012
0.044
0.012
–0.044
0.012
0.029
0.012
–0.029
0.012
0.019
0.012
–0.019
0.012
0.014
0.012
–0.014
0.012
0.008
0.012
–0.008
0.012
0.004
0.012
–0.004
0.012
0.002
0.012
–0.002
26
Ideal DC
Calibration
Output
Test Limit
(V)
(V)
–0.269
±0.000928
0.194
±0.000815
–0.194
±0.000815
0.144
±0.00074
–0.144
±0.00074
0.094
±0.000665
–0.094
±0.000665
0.064
±0.00062
–0.064
±0.00062
0.044
±0.00059
–0.044
±0.00059
0.029
±0.000567
–0.029
±0.000567
0.019
±0.000553
–0.019
±0.000553
0.014
±0.000545
–0.014
±0.000545
0.008
±0.000536
–0.008
±0.000536
0.004
±0.00053
–0.004
±0.00053
0.002
±0.000527
–0.002
±0.000527
Published
Specification
(V)
±0.003405
±0.00303
±0.00303
±0.00278
±0.00278
±0.00253
±0.00253
±0.00238
±0.00238
±0.00228
±0.00228
±0.002205
±0.002205
±0.002155
±0.002155
±0.00213
±0.00213
±0.0021
±0.0021
±0.00208
±0.00208
±0.00207
±0.00207