LabVIEW Block Diagram
NI 5402/5406 Calibration Procedure
9.
Close the instrument driver session and save the calibration date and
temperature by calling the niFgen Close Ext Cal VI.
Call
niFgen_CloseExtCal
using the following parameters:
vi: The session handle returned from
niFgen_InitExtCal
action: If the external adjustment
procedure completed without any
errors, use
. This function stores the new
COMMIT
calibration constants, updated
calibration dates, updated calibration
temperatures in the onboard
EEPROM.
If any errors occurred during the
external adjustment procedure, or if
you want to abort the operation, use
NIFGEN_VAL_EXT_CAL_
. This function discards
ABORT
the new calibration constants and
does not change any of the calibration
data stored in the onboard EEPROM.
94
C/C++
Function Call
NIFGEN_VAL_EXT_CAL_
ni.com