National Instruments NI 5402 Calibration Procedure page 86

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

LabVIEW Block Diagram
NI 5402/5406 Calibration Procedure
10. Initiate waveform generation by calling the niFgen Initiate
Generation VI.
11. Generate a sine wave configured for the current iteration of Table 10 at
each of the frequencies listed in Table 11. These steps will be repeated
for each iteration in Table 10.
Table 11. Frequencies for Adjusting Sine Wave Flatness Correction
Frequencies for the NI 5402
(Hz)
100,000
1,000,000
5,000,000
10,000,000
15,000,000
20,000,000
Frequencies for the NI 5406
86
C/C++
Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
(Hz)
100,000
1,000,000
5,000,000
10,000,000
15,000,000
20,000,000
25,000,000
30,000,000
40,000,000
.