8.
Close the instrument driver session, destroy the instrument driver
session and all of its properties, and release any memory resources
NI-FGEN uses by calling the niFgen Close VI.
LabVIEW Block Diagram
9.
If any of the errors are greater than the calibration test limits, perform
an external Adjustment.
13
C/C++ Function Call
Call
niFgen_close
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init
NI 5402/5406 Calibration Procedure