National Instruments NI 5402 Calibration Procedure page 55

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
Initializing Analog Output Calibration
1.
Connect the NI 5402/5406 CH 0 front panel connector to the DMM.
2.
Open an NI-FGEN external calibration session by calling the niFgen
Init Ext Cal VI.
LabVIEW Block Diagram
3.
Initialize analog output calibration by calling the niFgen Initialize
Analog Output Calibration VI.
LabVIEW Block Diagram
55
C/C++
Function Call
Call
niFgen_Init
using the
ExtCal
following parameters:
resourceName: The
name of the device that
you want to verify. This
name is the device
identifier assigned in
MAX.
password:
"NI"
C/C++
Function Call
Call
niFgen_
InitializeAnalog
OutputCalibration
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
NI 5402/5406 Calibration Procedure
.