National Instruments NI 5402 Calibration Procedure page 91

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

LabVIEW Block Diagram
© National Instruments Corporation
4.
Configure the waveform by calling the niFgen Configure Standard
Waveform VI.
LabVIEW Block Diagram
5.
Initiate waveform generation by calling the niFgen Initiate
Generation VI.
6.
Use the frequency counter to measure the frequency of the generated
waveform. This value is the first measured frequency that is used in
step 7b.
91
C/C++
Function Call
Call
niFgen_
ConfigureStandard
using the
Waveform
following parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
waveform:
NIFGEN_VAL_WFM_SINE
frequency:
10,000,000
amplitude:
1
startPhase:
0
dcOffset:
0
C/C++
Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
NI 5402/5406 Calibration Procedure
.
.