11. Determine the error for each measurement using the following
formula:
(Measured DC Output Value) – (Ideal DC Output Value)
12. Compare the DC error to the calibration test limits and published
specifications.
13. Abort waveform generation by calling the niFgen Abort
Generation VI.
LabVIEW Block Diagram
14. Close the instrument driver session, destroy the instrument driver
session and all of its properties, and release any memory resources
NI-FGEN uses by calling the niFgen Close VI.
LabVIEW Block Diagram
15. If any of the errors are greater than the calibration test limits, perform
a self-calibration procedure, followed by a reverification procedure.
DC Error =
37
C/C++ Function Call
Call
niFgen_Abort
Generation
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init
C/C++ Function Call
Call
niFgen_close
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init
NI 5402/5406 Calibration Procedure