NI 5402/5406 Calibration Procedure
30. Adjust the ADC calibration by calling the niFgen Cal Adjust Cal
ADC VI.
LabVIEW Block Diagram
31. Disable the analog output by calling the niFgen Property Node and
selecting Basic Operation»Output Enabled.
LabVIEW Block Diagram
52
C/C++
Function Call
Call
niFgen_Cal
AdjustCalADC
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
voltagesMeasured
Externally: {
external
,
measurement 0
external
}
measurement 1
voltagesMeasured
WithCalADC: {
cal
ADC measurement 0
cal ADC
}
measurement 1
C/C++
Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_OUTPUT_ENABLED
value:
VI_FALSE
.
,
.