National Instruments NI 5402 Calibration Procedure page 35

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
5.
Configure the waveform by calling the niFgen Configure Standard
Waveform VI.
LabVIEW Block Diagram
6.
Set the analog data mask by calling the niFgen Property Node and
selecting Output Attributes»Data Mask»Analog Data Mask.
LabVIEW Block Diagram
35
C/C++ Function Call
Call
niFgen_
ConfigureStandard
using the
Waveform
following parameters:
vi: The session handle
returned from
.
niFgen_init
channelName:
"0"
waveform:
NIFGEN_
VAL_WFM_SQUARE
frequency:
1,000,000
amplitude: The
Amplitude value for the
current iteration from
Table 6
startPhase:
0
dcOffset: The DC Offset
value for the current
iteration from Table 6
C/C++ Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init
channelName:
"0"
attributeID:
NIFGEN_ATTR_ANALOG_
DATA_MASK
value:
x0
NI 5402/5406 Calibration Procedure