National Instruments NI 5402 Calibration Procedure page 57

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
6.
Set the gain DAC value by calling the niFgen Property Node and
selecting Calibration»Gain DAC Value.
LabVIEW Block Diagram
7.
Set the offset DAC value by calling the niFgen Property Node and
selecting Calibration»Offset DAC Value.
LabVIEW Block Diagram
57
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_GAIN_DAC_VALUE
value:
2,000
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_ATTR_OFFSET_
DAC_VALUE
value:
32,767
NI 5402/5406 Calibration Procedure
.
.