National Instruments NI 5402 Calibration Procedure page 19

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

LabVIEW Block Diagram
© National Instruments Corporation
8.
Initiate waveform generation by calling the niFgen Initiate
Generation VI.
9.
Use the power meter to measure the output power level of the
NI 5402/5406. This value is the measured power of the 50 kHz sine
wave.
10. In this step you generate a sine wave configured for the current
iteration of Table 3 at each of the frequencies listed in Table 4.
Table 4. Frequencies for Verifying Sine Wave Flatness Correction Accuracy
Frequencies for the NI 5402
(MHz)
1.00000
5.00000
10.00000
15.00000
20.00000
25.00000
Frequencies for the NI 5406
19
C/C++
Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi: The session handle
returned from
.
niFgen_init
(MHz)
1.00000
5.00000
10.00000
15.00000
20.00000
25.00000
30.00000
40.00000
NI 5402/5406 Calibration Procedure