NI 5402/5406 Calibration Procedure
7.
The measured frequency should converge on the desired frequency.
If the measured frequency does not converge on the desired frequency
within 16 iterations, a problem may exist with your measurement
device or the NI 5402/5406.
a.
Abort waveform generation by calling the niFgen Abort
Generation VI.
LabVIEW Block Diagram
b.
Adjust the oscillator frequency by calling the niFgen Cal Adjust
Oscillator Frequency VI.
LabVIEW Block Diagram
92
C/C++
Function Call
Call
niFgen_Abort
Generation
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
C/C++
Function Call
Call
niFgen_CalAdjust
OscillatorFrequency
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
desiredFrequencyInHz:
10000000
measuredFrequencyIn
Hz: The measured
frequency value
(in Hz)
.
.