National Instruments NI 5402 Calibration Procedure page 73

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

© National Instruments Corporation
Adjusting the Main Path Post-Amplifier Gain and Offset
1.
Set the main DAC value by calling niFgen Write Binary 16 Analog
Static Value VI.
LabVIEW Block Diagram
2.
Set the waveform to user-defined by calling the niFgen Property Node
and selecting Standard Function Output»Waveform.
LabVIEW Block Diagram
73
C/C++
Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
value:
0
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
attributeID:
NIFGEN_
ATTR_FUNC_WAVEFORM
value:
NIFGEN_VAL_
WVFM_USER
NI 5402/5406 Calibration Procedure
.
.