National Instruments NI 5402 Calibration Procedure page 25

Hide thumbs Also See for NI 5402:
Table of Contents

Advertisement

Iteration
Digital Gain
1
1
2
–1
3
0
4
0
5
0
6
0
7
0
8
0
9
0
10
0
11
0
12
0
13
0
14
0
15
0
16
0
17
0
18
0
19
0
20
0
21
0
22
0
23
0
24
0
25
0
26
0
27
0
© National Instruments Corporation
Repeat steps 10 through 18 for each of the iterations listed in Table 5,
changing the Digital Gain, Amplitude, and DC Offset values for each
iteration.
Table 5. Values for Verifying DC Gain and Offset Accuracy
Amplitude
DC Offset
(V
)
pp
18
18
0.07
0.07
0.07
0.07
0.07
0.07
0.07
0.07
0.07
0.07
0.07
0.07
0.012
0.012
0.012
0.012
0.012
0.012
0.012
0.012
0.012
0.012
0.012
0.012
0.012
Ideal DC
Output
(V)
(V)
1
10
–1
–10
9.965
9.965
–9.965
–9.965
8.465
8.465
–8.465
–8.465
6.965
6.965
–6.965
–6.965
4.965
4.965
–4.965
–4.965
3.465
3.465
–3.465
–3.465
2.365
2.365
–2.365
–2.365
1.994
1.994
–1.994
–1.994
1.494
1.494
–1.494
–1.494
1.094
1.094
–1.094
–1.094
0.794
0.794
–0.794
–0.794
0.594
0.594
–0.594
–0.594
0.394
0.394
–0.394
–0.394
0.269
0.269
25
Calibration
Published
Test Limit
Specification
(V)
(V)
±0.038
±0.097
±0.038
±0.097
±0.015587
±0.052175
±0.015587
±0.052175
±0.013337
±0.044675
±0.013337
±0.044675
±0.011088
±0.037175
±0.011088
±0.037175
±0.008087
±0.027175
±0.008087
±0.027175
±0.005837
±0.019675
±0.005837
±0.019675
±0.004188
±0.014175
±0.004188
±0.014175
±0.003515
±0.01203
±0.003515
±0.01203
±0.002765
±0.00953
±0.002765
±0.00953
±0.002165
±0.00753
±0.002165
±0.00753
±0.001715
±0.00603
±0.001715
±0.00603
±0.001415
±0.00503
±0.001415
±0.00503
±0.001115
±0.00403
±0.001115
±0.00403
±0.000928
±0.003405
NI 5402/5406 Calibration Procedure