NI 5402/5406 Calibration Procedure
h.
Commit the attribute values to the device by calling the niFgen
Commit VI.
LabVIEW Block Diagram
i.
Wait 500 ms for the output to settle.
j.
Use the DMM to measure the NI 5402/5406 output voltage. This
measurement is
6.
Adjust the pre-amplifier main path gain and offset by calling the
niFgen Cal Adjust Main Path Pre Amp Gain VI.
LabVIEW Block Diagram
, which is used in step 6.
measurement 1
72
C/C++
Function Call
Call
niFgen_Commit
using the following
parameter:
vi: The session handle
returned from
niFgen_InitExtCal
C/C++
Function Call
Call
niFgen_Cal
AdjustMainPath
using the
PreAmpGain
following parameters:
vi: The session handle
returned from
niFgen_InitExtCal
channelName:
"0"
configuration: The
Current Configuration
value for the current
iteration from Table 8
mainDACValues:
{
, –
}
25233
29232
gainDACValues:
{
,
}
1500
2000
offsetDACValues:
{
}
32000
measuredOutputs:
{
,
measurement 0
}
measurement 1
.
.
ni.com