3 . P r i n c i p l e s o f O p e r a t i o n
3 . 2 : I n p u t a n d O u t p u t C o n n e c t i o n s
Input and Output Connections
The AutoCAT
®
2 Series will interface with most bedside monitors and can also receive
inputs directly from the patient. All input and output connectors are located on the front
of the pneumatic drive unit. There are two ECG input connectors (skin, High Level
monitor), three arterial pressure input connectors (FiberOptix
and High Level monitor).
Note: The FiberOptix
AP Signal requires the use of a CAL Key. This receptacle is
located next to the FiberOptix
FiberOptix
Series Arrow IAB catheter.
Three Patient Signal outputs are available for ECG, AP and BPW. The Assist Interval
output provides a signal output to a simulator for use during training or testing. Pump
information is available via the modem or RS232 serial output, and the Model 2001
simulator may be connected to the Simulator Connection.
Figure 3.8: AutoCAT
Inputs: Equipment such as patient monitors which can provide ECG and Arterial pressure
signal inputs to the IABP must be in compliance with IEC-601-1 and IEC-601-2-30.
• Patient simulators which provide ECG, AP and assist signal inputs used in
training and operational check-out of the AutoCAT
requirements.
Outputs: The AutoCAT
well as RS232 serial data to recorders, simulators, computers or bedside charting
systems. These devices must meet IEC60950 requirements.
A clear plastic cover is available (P/N 2800-92-64003) to cover and protect unused
I/O jacks. These may be easily inserted and removed from the pump.
Output Connections
3-14
sensor connector. The CAL Key is provided with each
Figure 3.7: Input and Output Connectors
2 WAVE
Fiber Optic Sensor, CAL key and balloon connector
®
®
Series can output ECG, AP and Balloon pressure signals as
sensor, transducer
Series must meet IEC60950