9 . O p e r a t i o n a l P r e c a u t i o n s , L i m i t a t i o n s a n d H a z a r d s
9 . 4 : S y s t e m M a l f u n c t i o n a n d P u m p S h u t d o w n
CAUTION
(page 10-6)
CAUTION
(page 10-9)
CAUTION
(page 10-11)
CAUTION
(page 10-16)
CAUTION
(page 11-16)
Explosion Hazard
(page 5-19)
Electric Shock Hazard
(pages 2-3 and 10-3)
System Malfunction and Pump Shutdown
9-10
Do not submerge electrical connectors during disinfection.
Secure a 0.3mm-thick polyethylene wrapping over the
connector before cleaning.
Do not use phenol-based cleaners. They cause cables to
harden and crack. Do not allow cables to remain immersed in
alcohol or other cleaning agents.
Visually inspect all cables and accessories including the ECG,
AP transducer/cable and power cord. If visible defects are
present, replace the cable or accessory. If a visible defect is
present in the power cord, DO NOT USE IN AC MODE.
Replace the power cord. Operate in Battery mode ONLY
until the power cord is replaced.
High pressure gas canisters should be handled by trained
personnel only.
Use only the fuse type and rating specified. Call Arrow
International's service number for assistance.
Panel covers should not be removed by anyone other than
Arrow International Field Service Engineers or their
authorized representatives. Shock hazards exist when the
protective covers are removed.
Do not turn dip switch 6 to the OFF position if Fiber optic
AP function is desired.
HELIUM USE ONLY label is located in the helium compartment.
TO REDUCE THE RISK OF ELECTRIC SHOCK, DO NOT
REMOVE COVER, REFER SERVICE TO QUALIFIED SERVICE
PERSONNEL label is located on the lower section of the left and
right side panels.
An explosion hazard exists with this system. Do not operate
the AutoCAT
®
2 Series IABP System in the presence of
flammable anesthetics or other gases.
An electric shock hazard may exist with this system. Always
operate the AutoCAT
2 Series from a 3-wire hospital-grade AC
®
electrical system with a separate ground. Do not remove the round
grounding pin from the system's plug. Do not use a 3-wire to 2-wire
adapter to avoid the system's ground. Do not place fluids in the
storage compartments on top of the AutoCAT
®
2 Series.