Signal Input Selection; Ecg Selection (Skin) With 5 Lead Ecg Cable - Arrow AutoCAT 2 Series Operating Manual

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

3 . P r i n c i p l e s o f O p e r a t i o n
3 . 3 : C o n t r o l K e y s a n d F u n c t i o n K e y s

Signal Input Selection

ECG Signal Source
AutoPilot
and Operator modes:
The ECG Signal source Control Key allows you to select the input source for the ECG.
There are two methods of connecting the ECG to the pump. These are:
1. ECG Skin using the Direct ECG Cable (5 Lead) or Backpad ECG. Select SKIN and
the desired lead. Available lead settings will appear in the multifunction keys, based
on the type of ECG cable connected to the AutoCAT
2. Phone to Phone (High Level Slave from bedside monitor). Select MONITOR under
ECG Select and verify the MONITOR LED is lit. ECG lead displayed on the
AutoCAT
®
2 Series will be the lead selected on the bedside monitor. When lead is
changed on the bedside monitor it will also change on the pump although MON is
displayed on the LCD (AutoCAT
this connection if the pacer output is available from the bedside monitor).
ECG Skin:
When the ECG Skin is selected and the ECG SELECT key is pressed, the multi-function
keys will change to reveal the lead selections, which are available to the pump. The 5-lead
cable has leads I, II, III, aVR, aVL, aVF and V available. When the 5 lead ECG cable is
in use, two ECG leads are displayed in the first 3 multi-function keys (see Figure 3.11).
When selecting leads for the 5-lead cable, the active lead is always displayed in the upper
portion of the multifunction key. If the desired selection is in the lower half of the
multifunction key, you must press the Multi-function key until the desired lead is in the
upper portion of the multi-function key and is highlighted in white. The lead displayed on
the LCD should match this selection.
I
II
AVR
AVL
F1
F2
Figure 3.11: ECG Selection (Skin) with 5 Lead ECG cable.

Signal Input Selection

3-28
2 WAVE
Series have pacer detection available via
®
®
III
V
AVF
MANUAL
F3
F4
2 Series.
®
AUTO
< GAIN
F5
F6
GAIN >
F7