AUTOMATIC RESPONSE ALARMS (Class II)
Problem
Operation Mode
ECG Lead Fault
AutoPilot
Detected
Operator
Trigger Loss
AutoPilot
8 . 2 : T r o u b l e s h o o t i n g A l a r m s C l a s s I I
Possible Cause(s)
CAL key not inserted or
corrupted.
FiberOptix
connector
needs to be replaced
or cleaned.
FiberOptix
electronics
failure.
FiberOptix
electronic
temperature out of range.
Altitude above 10,000 ft.
(5030 M)
Poor electrode connection
Loose connections
3 lead cable detected
Phono to Nicolay cable
detected
only
No ECG/AP/Pacer trigger
can be found
Very small ECG signal
8 . T r o u b l e s h o o t i n g
Corrective Actions
Insert CAL key. Change
IAB catheter. Use alternate
AP source.
Replace FiberOptix
connection. Clean
FiberOptix
sensor
connection point.
Call field service.
Replace console. Use an
alternate AP source. Call
field service.
Replace console. Use an
alternate AP source. Call
field service.
Use alternate AP source.
Re-apply electrode paste or
replace disposable electrodes
Check ECG cable
connections; repair/
reconnect as needed.
Replace ECG cable.
Use 5 lead ECG cables
only
Use a Phone to Phone cable
for slaving
Check patient condition
Switch to Operator mode
Check ECG/AP sources and
change as needed.
Use ECG gain to increase
ECG size.
Automatic Response Alarms
8-19