1 1 . P e r f o r m a n c e a n d T e c h n i c a l S p e c i f i c a t i o n s
1 1 . 1 : A u t o C A T
®
2 S e r i e s S p e c i f i c a t i o n s
AutoCAT
Inputs and Outputs
FiberOptix
Sensor
(AutoCAT
®
2 WAVE
®
only)
CAL Key
Technical Specifications
11-10
®
2 Series Technical Specifications
ECG MON input (high-level input):
± 5V full scale
accepts ECG signals from remote monitor
ART PRESS input:
calibrated at 100mmHg/V
accepts AP signal from remote monitor
ECG MON output (high-level output):
±3V full scale
provides ECG signal for display on remote monitor
ART PRESS output: 100mmHg/V
BALLOON PRESS output: 100mmHg/V
ASSIST INTERVAL:
TTL (used for interactive simulator connection)
ECG (patient cable input):
– for input of 5-lead patient cable
– maximum input 10 mV differential
ARTERIAL PRESSURE (transducer cable input):
compatible with any pressure transducer with output equivalent
to Spectramed transducer (50 µV/V/cm Hg)
BALLOON CONNECTOR:
electronic pins sense resistor value of IAB connector for input
of IAB size
DATA COMMUNICATIONS CHANNEL I:
DB-9 Connector (RS232) for serial transmission of
hemodynamic values, current alarms, time and date.
SIMULATOR:
DB-9 Connector (female) RS232 Model 2001 training simulator
provides AC power and patient signals to pump
MODEM:
for connection to a PC via phone line for remote system
monitoring
FLASH CARD
PCMCIA standard for storing data or downloading custom setups
For connection of Arrow IAB Catheter with Fiber Optic Sensor
Provides Fiber Optic Sensor information to IABP console
CAL Key provided with each Fiber Optic IAB