A p p e n d i x H : H e l p T e x t
KEY 1
KEY 2
HELP
ARR.
TIMING
ON/OFF
HELP
R-WAVE
DEFLATION
ON/OFF
KEY 1
KEY 2
HELP
HELP
INFLATION
TIMING
HELP
DEFLATION
TIMING
Appendix H
H-2
KEY 3
HELP MESSAGE TEXT
AUTOPILOT MODE (continued)
Turns automatic arrhythmia timing ON
or OFF in AUTOPILOT MODE ONLY.
When Arr. Timing is ON, the LED is lit
and the pump will automatically deflate
ON the R-Wave when conditions
permit. If OFF, deflation is predicted.
Available in Autopilot mode only. When
R-Wave deflation is turned ON the
pump will select AFIB trigger and
deflate on the R-Wave at all times,
whether arrhythmia is detected or not.
KEY 3
HELP MESSAGE TEXT
OPERATOR MODE
1. Select ECG/AP signals, select source
2. Connect balloon
3. Select trigger mode and assist ratio
4. Press PUMP ON
5. Adjust timing
INFLATION TIMING:
Set ASSIST RATIO to 1:2.
Locate DN between SYS and AUG
INFLATION: Set at, or just prior to
DN, so that: AUG > SYS
Check Deflation timing or set ASSIST
RATIO to 1:1
DEFLATION TIMING:
Set ASSIST RATIO to 1:2.
DEFLATION: Set deflation timing to:
ADIA < DIA & ASYS < SYS
Check Inflation timing or set
ASSIST RATIO to 1:1