3 . P r i n c i p l e s o f O p e r a t i o n
3 . 3 : C o n t r o l K e y s a n d F u n c t i o n K e y s
Alarm Message
Standby Alarm Disabled
To re-enable STANDBY alarm, press
Alarm Reset
PUMP On: Resume pumping
PUMP OFF: Stop pumping
Standby longer than 3 MIN
PUMP ON: Resume pumping
RESET: Continue Standby
STANDBY, STANDBY: Disable standby
Alarm indefinitely
ECG Trigger loss
(Operator mode only)
Pressure Trigger loss
(Operator mode only)
ECG Lead Fault detected
(Operator mode only)
Trigger Loss
1. ECG/AP/Pacer signals not found
2. Check Patient
3. Check ECG/AP connections
Alarm Messages
3-44
ALARM MESSAGES (Class II)
Description
3 minute Standby alarm suspended indefinitely.
Pump has been in standby for three minutes.
Eight seconds elapsed without a recognizable trigger point
in the ECG waveform (occurs only in PATTERN, PEAK,
A FIB, V PACER and A PACER trigger modes).
NOTE: ECG Trigger Loss Alarm is extended to
30 seconds when alarms are off.
Eight seconds elapsed without a recognizable trigger point
in the AP waveform (occurs only in ART PRESS trigger mode).
NOTE: Pressure Triggering Loss Alarm is extended to
30 seconds when alarms are off.
The AutoCAT
ECG leads (usually caused by loose or broken patient
leads). NOTE: The ECG lead which has the fault is shown
in the alarm message.
AutoPilot
mode only
No trigger can be established. Check patient condition.
®
2 detects high electrical impedance in the