ALARM MESSAGES (Class III) – Continued
Alarm Message
AP FOS Signal Weak
1. Check FOS connection
2. Use alternate AP source
3. Clean/service FOS connector
4. Call field service
AP FOS Sensor Out of Range
1. AP accuracy may be affected
2. Use alternate AP source
3. Call field service
AP FOS Cal key missing
or Corrupt
1. Connect AP FOS Cal key
2. Replace IAB catheter
3. Call Field Service
Alarm Messages
3-46
Description
AutoCAT
®
2 WAVE
®
only
Light source from AP FiberOptix
indicate a problem in the FiberOptix
FiberOptix
electronics or the connection point. Check
these components. Disconnect and reconnect the sensor.
Make sure a click is heard.
AutoCAT
®
2 WAVE
®
only
Electrical signal for AP FiberOptix
or is out of range. Switch to an alternate AP source.
AutoCAT
®
2 WAVE
®
only
FiberOptix
Cal key is disconnected or data has been lost.
Switch to an alternate AP source. Change IAB catheter if
FiberOptix
is needed.
is low. This may
sensor, the
cannot be detected,