Arrow AutoCAT 2 Series Operating Manual page 72

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

ALARM MESSAGES (Class III)
Alarm Message
Drain Failure
Excessive water build-up;
Repeat purge cycle
Possible drain valve failure
Deflation Timing beyond 100%
Timing Error
Insufficient time to Deflate
Check Trigger mode
Select Operator mode
Change trigger/timing settings
Insufficient Time to Inflate
Warning: Battery Inoperative
Call Field Service
Available Battery Power
Less than 5 minutes
Available Battery Power
Less than 10 minutes
Available Battery Power
Less than 20 minutes
System Running on Battery Power
Possible ECG Trigger Detected
Weaning Step Complete
Evaluate Hemodynamics, set
Parameters, press START WEANING
To initiate next step.
Arterial Pressure Alarm
AP has fallen below set limit:
1. Check AP for disconnect
2. Assess patient hemodynamics
3. Change AP alarm limit
Low Helium Tank Pressure
3 . P r i n c i p l e s o f O p e r a t i o n
3 . 3 : C o n t r o l K e y s a n d F u n c t i o n K e y s
Description
Cannot remove condensate
Operator mode only
Check for proper timing
Inflation and deflation points need to be adjusted
(insufficient time to deflate the balloon before the next
inflation cycle).
Inflation and deflation points need to be adjusted. There
is not enough time to inflate before deflation is set to
occur. If in AFIB timing and Operator mode, move
inflation earlier.
The AutoCAT
®
2 will not run in battery mode due to a
faulty DC circuit breaker or circuit breaker is OFF.
Less than 5 minutes of battery power remain before
system battery operation shuts down.
Less than 10 minutes of battery power remain before
system battery operation shuts down.
Less than 20 minutes of battery power remain before
system battery operation shuts down.
AC power was intentionally or accidentally disconnected
and the AutoCAT
2 has automatically switched to
®
battery power.
Operator mode only
Internal trigger selected and ECG signal is available.
Check patient condition and select an ECG trigger mode.
Weaning Timer Expired
AP has fallen below set limit
Helium tank pressure < 100 psi
Alarm Messages
3-45