5 . O p e r a t i n g P r o c e d u r e s
5 . 1 : O p e r a t i n g I n s t r u c t i o n s
ACTION
4. Connect AP source
5. Connect IAB
6. Verify reliable trigger mode • Verify Heart Rate
7. Initiate Counterpulsation
8. Perform timing assessment • Verify automatic timing
9. Check BPW
10. Verify alarms are ON
Summary
5-4
AUTOPILOT MODE
DETAILS
• AP FiberOptix
WAVE
• Connect the FiberOptix
key to the IABP
• Zero the FiberOptix
insertion of the IAB
and/or
• AP TRANSDUCER:
and/or
• AP MONITOR:
• Verify waveform appears on display.
• Check AP scaling
• Volume automatically set to full volume
• Verify white bands on ECG
• Check Trigger mode displayed below HR
• Press Pump ON
• Verify IAB size comparing AUG to BPW
plateau pressure
• Verify BPW shape
sensor: AutoCAT
2
®
only
®
sensor and CAL
sensor PRIOR to