Arrow AutoCAT 2 Series Operating Manual page 259

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

Table 206
Recommended separation distances between
portable and mobile RF communications equipment and the AutoCAT
The AutoCAT
®
2 Series is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the AutoCAT
Series can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the AutoCAT
®
2 Series as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum Separation distance according to frequency of transmitter
output power of
transmitter
w
150 kHz to 80
MHz
d = 1.2√P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where p is the maximum output power rating of
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
A p p e n d i x : A p p l i c a b l e S t a n d a r d s
m
80 MHz to 800
800 MHz to 2.5 GHz
MHz
d = 2.3√P
d = 1.2√P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
®
2 Series
2
®
Appendix: Applicable Standards
7

Advertisement

Table of Contents
loading

Related Products for Arrow AutoCAT 2 Series

This manual is also suitable for:

Autocat 2Autocat 2 wave

Table of Contents