3 . P r i n c i p l e s o f O p e r a t i o n
3 . 3 : C o n t r o l K e y s a n d F u n c t i o n K e y s

Balloon Volume

AutoPilot
and Operator Modes:
The AutoCAT
2 SERIES automatically sets the volume from the IAB connector. However
®
when volume changes are required the AutoCAT
®
2 Series allows the user to set the precise
volume to be delivered to the IAB in 0.5 cc increments. Volume can be changed while the
pump is OFF or during pumping. If IAB volume is changed while pumping, the pump will
pause for 1 or 2 beats to reset the volume and then resume pumping at the new volume
setting.
When the BALLOON VOLUME function key is pressed, the multi-function keys will
change to the following:
DELIVERY VOLUME: 40cc 100%
<
>
FULL
DECREASE
INCREASE
VOLUME
F1
F2
F3
F4
F5
F6
F7
Figure 3.12: Initial display of volume change.
DELIVERY VOLUME: 32cc 80%
<
>
FULL
CANCEL
APPLY
DECREASE
INCREASE
VOLUME
F1
F2
F3
F4
F5
F6
F7
Figure 3.13: Example of IAB volume change to less than full volume.