Trigger Selection In Autopilot; Mode - Arrow AutoCAT 2 Series Operating Manual

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

Trigger Selection in AutoPilot

When Autopilot mode is selected, the pump will choose the trigger mode based on
information available from the patient and on the IABP settings selected by the user.
The following is the trigger mode selection criteria in AutoPilot
Trigger mode
PATTERN (Default)
PEAK
AFIB
VPACE
APACE
AP
NOTE: Arrhythmia detection criteria: 8 of the last 16 beats vary more than 15% on a
beat to beat basis.
NOTE: APACE trigger may be selected when an ECG or AP signal are present but are
not yet stable.
NOTE: In the event that there are more than 3 trigger switches between Peak and
Pattern within 1 minute, the pump will automatically select PEAK trigger for 3 minutes.
This allows timing and triggering to be more stable during periods of significant heart
rate changes and/or an unstable ECG signal. This selection will be indicated by PEAK
trigger being displayed in light Blue in the trigger area of the display. In the event that
Peak trigger becomes unavailable or an arrhythmia is detected the pump will
automatically select the appropriate trigger mode based on current patient conditions.
4 . P r i n c i p l e s o f T i m i n g a n d T r i g g e r i n g
4 . 3 . 1 : T r i g g e r S e l e c t i o n i n A u t o P i l o t

Mode:

AUTOPILOT MODE SELECTED
Criteria
HR < 130 bpm
No arrhythmia detected
HR > 130 bpm
Arrhythmia detected and Arrhythmia timing OFF
Arrhythmia detected and Arrhythmia timing ON but
conditions are not appropriate for R-Wave deflation.
HR: any
Arrhythmia detected, Arrhythmia Timing On and
conditions appropriate for R-Wave deflation.
R-Wave deflation ON
No ECG or AP signal present
Single Pacer with no ECG
Dual pacer (A and V spike < 250 msec apart)
Single pacer with ECG present and time of
> 100 msec from pacer upstroke to R wave
No ECG signal available
Noisy ECG signal
Bovie detected noise
M o d e
:
Trigger Selection in AutoPilot
Mode
4-5