8 . T r o u b l e s h o o t i n g
8 . 1 : C o m m o n O p e r a t i o n a l P r o b l e m s
Section 1 Troubleshooting the FiberOptix
The following sections provide detailed information on troubleshooting issues with the
Fiber Optic system. Subsequent pages list the status messages that may be viewed by
pressing HOME then SHOW STATS. These codes provide the current operational status
of the FOS electronics, connections and sensor. A description of the code, possible
alarms associated with that status code and the issues which may result in the code being
displayed are shown. Subsequent pages show possible symptoms, the possible related
status codes, possible causes and recommended troubleshooting actions.
NOTE: In all cases, if the Fiber optic system is or becomes non-operational, an alternate
AP source will be selected by Autopilot. If Operator mode is selected, an alternate AP
source should be selected by the user.
NOTE: The loss of the fiber optic signal does not impair functionality of the IABP system.
Section 2 Fiber Optic Status Codes
This section details the status messages that are issued by the Fiber optic system. They
are designed to verify the current FOS status and assist in providing specific information
for troubleshooting FOS issues.
Status Code
Description
OK
FOS electronics and
sensor are working
properly and ready
for use.
NO COMM
No communication
between CPU and
FOS electronics
LL
Low light return
SL
Strong light
CK
Cal key missing or
data is corrupt
Troubleshooting the AP FOS signal
8-6
System
FOS Status Codes
Possible Alarm
None
System Error 8
AP FOS Weak Signal
AP FOS Signal
Out of Range
AP FOS CAL Key
Missing or Corrupt
Possible issues
*None
*FOS PCB failure
*FOS PCB or CPU cable
disconnect
*Loss of power to FOS PCB
*CPU failure
*Sensor not fully connected
*FOS cable is damaged or
broken
*Connections are dirty
*Connection is damaged
*FOS board failure
*Sensor not properly connected
*Connections are dirty
*Connection is damaged
*CAL key not connected
*CAL key data corrupt Ribbon
cable loose or disconnected

Advertisement