Arrow AutoCAT 2 Series Operating Manual page 251

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

KEY 1
KEY 2
MESSAGES WHICH ARE THE SAME IN AUTOPILOT AND OPERATOR MODES
HELP
ALARMS
ON/OFF
HELP
RECORDER
ON/OFF
HELP
BALLOON
VOLUME
HOME
HELP
HOME
HELP
HOME
HELP
KEY 3
HELP MESSAGE AND TEXT
Turns alarms audio, recording, drain
and refill on or off. To select alarm
time off, press key under desired
setting. Alarm messages will still be
displayed. Time remaining For alarms
off is displayed above the AP Scale.
Press again to turn on alarms.
Starts and stops recorder. To change
recorder settings press HOME and
RECORDER SETUP
Select INCREASE/DECREASE until
desired volume is displayed. Press
FULL VOLUME to return to volume
based on balloon connector. Press
APPLY to change volume or Press
CANCEL to cancel changes made.
AP SCALING
Press AP SCALING to select Auto or,
Manual scaling. Auto ON selects AP
scale which displays the entire
waveform. Autoscaling OFF allows
user to select AP scale which does not
change. Current selection is
highlighted.
RECORDER
Press RECORDER SETUP to show
SETUP
choices. Select one or two waveforms,
change speed and set timed recording.
Choices are highlighted. To change,
press key under selection. Waveforms
cannot be changed while recording.
WEANING
WEANING SETUP sets volume,
SETUP
assist ratio and time for a weaning
session. Select parameter and change
using the < and > keys. Press START to
begin weaning. A timer is displayed on
the main screen. Press the 100% Vol @
1:1 key X 2 to cancel Weaning and
resume full support.
A p p e n d i x H : H e l p T e x t
Appendix H
H-5