Fuse Replacement

The AutoCAT
®
2 Series has two fuses located in the pump module. These should be
changed only by Arrow International Field Service Engineers or trained personnel.
See Section 10.3 for ordering information. Part numbers for the fuses can be found in
Section 10.3.
Use only the fuse type and rating specified. Call Arrow International's service number for assistance.
Panel covers should not be removed by anyone other than Arrow International Field Service Engineers
or their authorized representatives. Shock hazards exist when the protective covers are removed.
To change the fuses (if necessary), remove the AC power cord as described in Section
2.1, use a small flat blade screwdriver to press on the fuse holder locking lever and
remove the fuse drawer which carries two fuses. Remove fuses from the drawer, insert
new fuses into the drawer and install fuse drawer back into place. Press the drawer in
until it clicks indicating that it has locked into position. Reinstall the power cord as
described in Section 2.1.

System Shutdown

If the AutoCAT
®
2 Series will be idle for four weeks or more, shut down the system as
follows:
1. Press the power switch to turn off the power.
2. Remove the disposable helium tank or turn off disposable/refillable tank (see
instructions in this section).
3. Empty the condensate bottle (see instructions in this section).
4. Store all required cables with the console.
Note: Leave the power cord connected to an AC power supply to maintain a full
charge on the battery. The system automatically prevents overcharging.
1 0 . M a i n t e n a n c e a n d S e r v i c e
1 0 . 1 : R o u t i n e M a i n t e n a n c e P r o c e d u r e s
CAUTION
CAUTION

System Shutdown

10-11