9 . O p e r a t i o n a l P r e c a u t i o n s , L i m i t a t i o n s a n d H a z a r d s
9 . 3 : H a z a r d s A s s o c i a t e d w i t h t h e A u t o C A T
Transport may involve high noise environments. In some
environments, audible alarms may not be heard by the
operator. Therefore, it is highly recommended that the
operator have clear visibility of the LCD in transport situations
since alarms are displayed on the LCD when they occur.
The FiberOptix™ sensor must be zeroed prior to placing the
IAB in the patient. Failure to zero the transducer may result
in inaccurate AP values. The non-zeroed FiberOptix
signal will be adequate for WAVE
WAVE
®
but should not be used for hemodynamic assessment
of the patient.
Leave the CAL key connected at all times. If the CAL key is
removed but the same CAL key is reinserted, the calibration
and zero values are retained. If another CAL key is inserted
all previous cal and zero values will be lost and the AP
FiberOptix
sensor readings accuracy may be affected.
Use only the CAL key supplied with the IAB. Each
FiberOptix™ sensor has a unique CAL key. Use of another
CAL key may affect the accuracy of the FiberOptix™ sensor
AP readings.
Verify that the AP MAP value that is used to calibrate the AP
FOS MAP is zeroed and calibrated. The AP source should be
verified for accuracy as well.
Follow the manufacturer's instructions to connect the
heparinized fluid source to the constant flush unit. All
pressure tubing, stopcocks and connections must be filled
completely with fluid and be free from air bubbles before
proceeding with calibration.
Panel covers should not be removed by anyone other than
Arrow International Field Service Engineers or their
authorized representatives. Shock hazards exist when the
protective covers are removed.
Do not clean the AutoCAT
patient.
Clean and disinfect using only those solvents listed. Do not
use other solvents. Avoid acetone, 100% phenol cleaners,
ether or higher concentrations of formaldehyde. These
chemicals can damage the console's finish and accessories.
Examine the cable's outer casing carefully for perforations
before cleaning. Do not soak a perforated cable. Have it
replaced immediately.
timing in the AutoCAT
®
®
2 Series while it is connected to a
Hazards Associated with the AutoCAT
®
2 S e r i e s
sensor
2
®
®
2 Series
9-9