Arrow AutoCAT 2 Series Operating Manual page 65

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

3 . P r i n c i p l e s o f O p e r a t i o n
3 . 3 : C o n t r o l K e y s a n d F u n c t i o n K e y s
If IABP volume is changed while pumping patient support will be momentarily suspended as the volume is
updated. Insure the patient will tolerate this procedure before pressing APPLY to initiate the volume change.
Change Volume:
1. Press BALLOON VOLUME key.
2. INCREASE or DECREASE the volume to the desired setting.
3. Press APPLY to change volume.
NOTE: Pump will reset volume in 1 or 2 beats and resume pumping.
4. If volume change was made in error, press CANCEL or wait 30 seconds for
multifunction keys to time out.
Return to Full Volume:
1. Press BALLOON VOLUME key.
2. Press FULL VOLUME multifunction key.
3. Press APPLY to change volume.
NOTE: Pump will reset volume in 1 or 2 beats and resume pumping. If volume change
was made in error, press CANCEL or wait 30 seconds for multifunction keys to time out.
NOTE: If the volume is changed from the current delivered volume and then returned to
that volume, the CANCEL and APPLY keys will disappear. These keys are only
available when changing the volume from the current delivered volume setting.
NOTE: If the pump is in Stand-by when the volume change is made, the pump will go to
OFF when APPLY is pressed. Press pump ON to start pumping at the new volume
setting.
Balloon Volume
3-38