A p p e n d i x : A p p l i c a b l e S t a n d a r d s
Conformance to Standards:
STANDARD
CISPR 11 EN 55011
IEC 61000-3-2
IEC 61000-3-3
IEC 61000-4-2
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-11
IEC 61000-4-8
IEC 61000-4-6
IEC 61000-4-3
Table 201
Guidance and manufacturer's declaration - electromagnetic emissions
AutoCAT
®
2 Series IABP is intended for use in the electromagnetic environment
specified below. The customer or the user of the AutoCAT
assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Harmonic
Not
emissions
applicable
IEC 61000-3-2
Voltage
Not
fluctuations/
applicable
flicker emissions
IEC 61000-3-3
Appendix: Applicable Standards
2
Table # (referenced to IEC 60601-1-2)
Table 201
Table 201
Table 201
Table 202
Table 202
Table 202
Table 202
Table 202
Tables 204, 206
Tables 204, 206
®
Electromagnetic environment - guidance
The AutoCAT
®
2 Series IABP uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The AutoCAT
®
2 Series IABP is suitable for use
in all establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
2 Series IABP should