Arrow AutoCAT 2 Series Operating Manual page 249

Intra-aortic balloon pump iabp system
Table of Contents

Advertisement

KEY 1
KEY 2
HELP
TRIGGER
MODE
TRIG-
HELP
GER
MODE
TRIG-
HELP
GER
MODE
TRIG-
HELP
GER
MODE
TRIG-
HELP
GER
MODE
TRIG-
HELP
GER
MODE
KEY 3
HELP MESSAGE TEXT
OPERATOR MODE (continued)
Selects Trigger in Operator mode
Automatic trigger selection in
AutoPilot
select OPERATOR, then press Trigger
mode. Trigger mode is selected by
pressing the key under the desired
trigger mode.
PATTERN
PATTERN trigger mode.
Preset trigger; for normal QRS
complex. Uses height, width, and slope
of positive or negative QRS complexes.
Width must be between 25 and 135
mSec. Rejects pacer spikes.
PEAK
PEAK trigger mode
For any type of QRS complex and
changing QRS shapes. Uses height,
slope only of positive or negative QRS
complexes. May be preferred for HR >
140. Rejects pacer spikes.
AFIB
AFIB trigger mode
For irregular cardiac rhythms. Uses
height, slope only of positive or
negative QRS complexes with REAL
TIME (R-Wave) deflation. Rejects
pacer spikes.
VPACE
VPACE trigger mode
Uses V-pacer spikes to trigger, MUST
BE 100% PACED. For V and AV
sequential pacers. ECG SKIN cable
connection recommended.
APACE
APACE trigger mode
Uses A-pacer spikes to trigger, MUST
BE 100% PACED. For Atrial pacers
only. ECG SKIN cable connection
recommended.
A p p e n d i x H : H e l p T e x t
. To change trigger mode,
Appendix H
H-3

Advertisement

Table of Contents
loading

Related Products for Arrow AutoCAT 2 Series