AutoCAT
Strip Chart
Recorder
(cont.)
Operating Gas
Helium Tank
Volume/Pressure
Control
Water Vapor
Removal
Gas Drive
Pumping Volume
ECG Low Level
Filtering
Polarity
ECG Low Level
Bandwidth
Leakage Current
Line Isolation
Defibrillator
Protection-ECG
Defibrillator
Protection-Arterial
Pressure
1 1 . P e r f o r m a n c e a n d T e c h n i c a l S p e c i f i c a t i o n s
1 1 . 1 : A u t o C A T
®
2 Series Technical Specifications
– assist ratio
– IAB volume delivered/ECG lead/Timing Settings/Trigger
Modes/Operation Mode/AP Alarm Status/Timing Method/Assisted
and Unassisted AP values
– Automatic recordings of Class 1 Alarms
Speed: 25mm/sec. (±5% of rated speed)
Paper: 50mm (±0.03mm)-wide, blank thermal paper
(Roll Diameter not to exceed 5.4cm)
Resolution: 400 dots/inch @ 25mm/sec.
USP helium
WARNING:
Do not use oxygen or any drive gas other than
USP helium.
Disposable 500psi canister or
2000psi refillable cylinder
Closed loop system
Solenoid-actuated, thermo-electric baffle system removes moisture
from pneumatic lines. Collection bottle can be emptied without
interrupting operation.
Stepper motor-driven bellows (helium gas drive only)
0-50 cc in 0.5 cc increments
Diathermy detection, 30Hz Low pass
Automatic processing of positive or negative triggering signals
(Arterial Pressure must be positive)
0.75-30Hz
Less than 10 µA
120 db at 60Hz to ground
ECG input protected up to 400 joule, 5 kV peak defibrillator
discharges at 20-sec intervals. Meets IEC-60601-2-25
Fiber optic signal is non-conductive.
AP Transducer: Meets IEC-60601-2-34 (applicable sections only)
®
2 S e r i e s S p e c i f i c a t i o n s
Technical Specifications
11-9