A p p e n d i x : A p p l i c a b l e S t a n d a r d s
Table 204 (continued)
Guidance and manufacturer's declaration - electromagnetic immunity
Immunity test
IEC 60601
test level
Radiated RF
80 MHz to
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
ª Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which
the AutoCAT
®
2 Series is used exceeds the applicable RF compliance level above, the
AutoCAT
®
2 Series should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting
or relocating the AutoCAT
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Appendix: Applicable Standards
6
Compliance
level
2 Series.
®
Electromagnetic environment -
guidance
d = 2.3√P 800 MHz to 2.5 GHz
Interference may occur in the
vicinity of equipment marked
with the following symbol: