A p p e n d i x H : H e l p T e x t
KEY 1
KEY 2
TRIG-
HELP
GER
MODE
TRIG-
HELP
GER
MODE
KEY 1
KEY 2
MESSAGES WHICH ARE THE SAME IN AUTOPILOT AND OPERATOR MODES
HELP
ECG
SELECT
HELP
AP SELECT
HELP
CURSOR
Appendix H
H-4
KEY 3
HELP MESSAGE AND TEXT
OPERATOR MODE (continued)
ARTERIAL
AP trigger mode
PRESSURE (AP) Uses AP waveform to trigger.
Recommended when ECG is not
available or too noisy. NOT
RECOMMENDED FOR IRREGULAR
RHYTHMS.
INTERNAL
INTERNAL Trigger mode:
Uses IABP internal signal for
triggering. Used when no ECG or
APsignal is available. ASYNCHRONOUS
TO PATIENT CARDIAC ACTIVITY.
Press INT again to confirm.
KEY 3
HELP MESSAGE AND TEXT
ECG sources are automatically selected
in AUTOPILOT. User can change
LEAD, source, gain mode and level. To
change source, press ECG SELECT
again. LEAD label. To switch gain mode
press key under desired label. Use <
and > keys to adjust manual gain.
AP SELECT provides selection for
source SCALE, AP Alarm, ZERO and
CAL. To change input source, press AP
SELECT again. Press AP SCALING
for scaling. To zero, open transducer to
air and press ZERO. To CAL, input
100mmHg and adjust sens. AP alarm:
MAP/AUG, press ON, set limit.
Moves horizontal cursor on AP and
BPW. Move cursor to desired
assessment point. Value is displayed
above cursor on the right hand side.